84: krypton, Site willow, Tue 17-Aug-2004 15:19:39 MDT, Cleaned 2m Kh20
~13:45 local
Distilled water used.
Note began raining within 45min or so afterwards so
the data may not show improved signal for a bit.

81: krypton, Site aspen, Tue 17-Aug-2004 13:00:32 MDT, Cleaned 2m Kh20
12:25 local
Distilled water used.
Note began raining within 15min or so afterwards so
the data may not show improved signal for a bit.
73: krypton, Site pine, Thu 12-Aug-2004 15:11:15 MDT, krypton test/clean
from 21:05-2108 I was playing with getting a better krypton signal -- I failed.

69: krypton, Site all, Tue 10-Aug-2004 17:46:43 MDT, All 1M Krypton Cleaned
08/10/04
All 1-m kryptons were cleaned with distilled h2o

Pine:   12:30 local, 19:30z
Willow: 14:10 local, 21:10z  Note: cobwebs on windows
Aspen:  14:45 local, 21:45z

63: krypton, Site all, Mon 09-Aug-2004 13:32:24 MDT, all kryptons "cleaned"
I just finished a round of cleaning all kryptons with alcohol.  The order was:
30m.a
2m.a
2m.p
2m.w
17m.w
30m.p
However, the voltages for all of the lower ones are still quite low (0.04V at
2m.p).  I'll probably go back to work on this one.

25: krypton, Site willow, Tue 06-Jul-2004 14:55:41 MDT, 2m krypton at willow
The 2m krypton appears to be ok. I checked it with a DVM. The problem is either
the serializer or the connection to the the serializer. Unfortunately the
test cable with the DVM broke so I was not able to continue tests. Also the
adapter cable from the krypton to the sonic had not been potted. Water was in
the connector.